▷ Mount the push-in angle piece of the supplied
gas hose to the stop onto the hose piece of the
sintering base, until the resistance of the seal
has been overcome. When doing this, counter-
hold the sintering base in the furnace compart-
ment.
A click noise will indicate that the push-in con-
nection is properly engaged.
Fig. 16
If connecting the Argovent 2 fails:
▷ Remove the sintering base.
▷ Outside of the furnace, hold the sinter-
ing base by the connection pin and care-
fully mount the gas hose onto the hose
piece of the sintering base.
This slightly expands the seal and makes
connecting the Argovent in the furnace
easier.
▷ Undo the connection again, place the
sintering base into the furnace and re-
establish the connection.
I N S T A L L A T I O N
▷ Mount the back panel of the furnace and screw
in the six screws.
Fig. 17
▷ Mount the insulating stone, ensuring that the
ceramic studs engage into the holes in front of
the floor plate.
Fig. 18
To check if the insulating stone is properly seated:
▷ Carefully close the furnace lift door.
▷ Insert mains plug again.
E N