CONSERVATION AND CORRECTIVE AND PREVENTIVE
MAINTENANCE
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing
to be cleaned and disinfected.
Cleaning
Important: In order to execute cleaning or any type of maintenance, ensure that the
equipment is disconnected from the electrical network.
The cleaning procedure should be executed at the start of the
working day and after each patient. Always turn the main switch
off before executing the procedures of daily maintenance.
• BODY OF THE EQUIPMENT, CABLE, TRANSDUCTOR AND HOSES
To clean the equipment, we recommend the use of "BactSpray" (Reg nº
MS: 3.2079.0041.001-5) or any other similar product: Active component:
Benzalkonium chloride (tri-quaternary ammonium)
Solution 50%................................................. 0.329%
Chemical composition: Butyl Glycol, Decyl polyglucose, Sodium Benzoate,
Sodium Nitrate, Essence, Deodorized Propane / Butane, demineralized Water.
For more information concerning cleaning procedures, see manufacturer's
instructions.
NOTE: The registration at the Ministry of Health of the "BactSpray"
is executed separately from the product described in this manual, as the
"BactSpray" is not manufactured by Gnatus.
WARNING:
- In order to prevent risks and damages to equipment, make sure that the liquid does
not enter into the equipment.
- The application of other solvent-based cleaning products or sodium hypochloride isn't
recommended, because they may damage the equipment.
Disinfection
- Use clean and soft cloth dampened in alcohol 70% to disinfection of the equipment.
- Never use corrosive disinfectants or solvents.
Note: Use gloves and other
systems of protection,
during the disinfection.
48