Technical Specifications; General; Characteristics; Dc Voltage (Autorange) - Metrix MX 350 Manual De Instrucciones

Ocultar thumbs Ver también para MX 350:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5.

TECHNICAL SPECIFICATIONS

5.1. General

Only the values assigned tolerances or the limits
declared constitute guaranteed values. The values
without any tolerance are given as indications.

5.2. Characteristics

The accuracy is ± [% of the reading (R) + number of
representation units (digits or D)] in the reference
conditions (see appendix).

5.2.1. DC voltage (Autorange)

Range
Measurement range
Resolution
400 V
0.2 V to 399.9 V
0.1 V
600 V
400 V to 600 V
1 V
Protection against overloads: 660 Vrms

5.2.2. AC voltage (Autorange)

Range
Meas. range
Frequency
400 V
0.5 V to 399.9V 50 .. 500 Hz
600 V
400 V to 600 V
50 .. 500 Hz
MX 350: Input impedance: 1 MΩ
MX 355: Input impedance: 10 MΩ
Protection against overloads: 660 Vrms

5.2.3. DC current (Autorange)

(MX 355)
Range
Measurement range
Resolution
40 A
0.10 A to 39.99 A
0.01 A
400 A
40.0 A to 400.0 A
0.1 A
Protection against overloads: 600 Arms

5.2.4. AC current (Autorange)

(MX 350)
Range
Meas. range
Frequency
40 A
0.05A to 39.99A 50 .. 60 Hz
60 ... 500 Hz
400 A
40.0A to 400.0A 50 .. 60 Hz
60 ... 500 Hz
Protection against overloads: 600 Arms
(MX 355)
Range
Meas. range
Frequency
40 A
0.05A to 39.99A 50 .. 500 Hz
400 A
40.0A to 400.0A 50 .. 500 Hz
Protection against overloads: 600 Arms
28
Chapter II
Accuracy
1% R +2 D
1% R +2 D
Resol.
Accuracy
0.1 V
1.5%R +5 D
1 V
1.5%R +5 D
Accuracy
2.5% R +10 D
2.5% R +10 D
Resol.
Accuracy
0.01 A
1.9%R + 5 D
2.5%R + 5 D
0.1 A
1.9%R + 5 D
2.5%R + 5 D
Resol.
Accuracy
0.01 A
2%R +10 D
0.1 A
2%R +10 D
MX 350 / MX 355

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mx 355