ResMed S9 Serie Guia De Informacion página 6

Ocultar thumbs Ver también para S9 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

 
Problem/possible cause
Air feels too warm/cold in the mask
The temperature of the ClimateLine air
tubing is set too high/low.
General technical specifications
Power supply
90W
unit
30W
Environmental
Operating temperature: 41°F to 95°F (+5°C to +35°C)
conditions
Operating humidity: 10-95% non-condensing
Operating altitude: Sea level to 8,500' (2,591m)
Storage and transport temperature: -4°F to 140°F (-20°C to +60°C)
Storage and transport humidity: 10-95% non-condensing
Aircraft use
ResMed confirms that this device meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160F, section 21, category M) for all phases of air
travel.
Electromagnetic
Product complies with all applicable electromagnetic compatibility requirements
compatibility
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. Information regarding the electromagnetic emissions and
immunity of this ResMed device can be found on www.resmed.com, on the
Products page under Service and Support. Click on the PDF file for your
language.
IEC 60601-1
Class II (double insulation), Type BF
classification
S9 technical specifications
Pressure and
Operating pressure range: 3 to 25 cm H
flow state
Maximum single fault steady state pressure: 40 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±6 L/min or 10% of reading, whichever is greater,
at 0 to 150 L/min positive flow
6
Solution
Turn up/down the heated tubing temperature via the menu.
Input range: 100-240V, 50-60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
Input range: 100-240V, 50-60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 20W (30VA)
Maximum power consumption: 36W (75VA)
O
2
O
2
O ± 4% of the measured reading
2

Publicidad

Tabla de contenido
loading

Tabla de contenido