Step 6.
Measure samples
Put the sample to be measured in the main light path and put
Reference in the reference light path, then the result displays on
the screen automatically.
9. Quantitation
Step 1.
Start Quantitation
Main menu, press
followed with
WL: 500.0nm
Main Menu
Basic Mode
①
Quantitative Mode
②
Wavelength Scan
③
Kinetics Mode
④
DNA/Protein Mode
⑤
Multi-Wavelength Mode
⑥
System Utility
⑦
Cancel
Step 2.
Set Unit
Press
to set concentration unit, press
EN TER
press
to confirm.
Step 3.
Establish Standard Curve or load the stored curve
Press
to go into set up interface, 2 methods are under your
choice.
WL: 500.0nm
Quantitative Test
No.
Abs
C=0.226+1.031*A+^1
Unit
Fit Curve
Establish Standard Curve:
Method 1: Input Regression Equation
4) Set Fit Curve Method. Press
to choose the method and press
2
ABC
or press
EN TER
pressed to confirm.
10:00:00
D2
W
Move Up
Move Dn
Select
10:00:00
D2
W
Conc.(mg/ml)
WL.(nm)
500.0
Search
Scroll
r=0.962
,
to choose "Quantitative Mode"
WL: 500.0nm
Quantitative Test
No.
Abs
Conc.(mg/ml)
C=0.226+1.031*A+^1
r=0.962
Unit
Fit Curve
,
WL: 500.0nm
Calibration Table
No.
Conc.(mg/ml)
1
0.000
C=0.226+1.031*A+^1
r=0.962
Method
Params
to set Fit method,use
EN TER
to confirm.
10:00:00
D2
W
WL.(nm)
500.0
Search
Scroll
to choose and
10:00:00
D2
W
Abs
WL.(nm)
500.0
Standard
Show Curve
,