Idiomas disponibles

Idiomas disponibles

5) Set Wavelength. Press
choose measure method, then press
wavelength value you need and press
6) Input the Factor of the Regression Equation. Press
input the factors, press
Method 2: Use Standard Samples
5) Set Fit Method. Press
choose fit method, then press
6) Set
Wavelength.
interface, press
EN TER
to confirm. Input the value of the wavelength and press
EN TER
to confirm.
7) Setup Standard Samples. Press
the concentrations of corresponding standard samples according
the indication and press
to choose the value you just input and press
then input a new value, press
cancel after all the input.
WL: 500.0nm
Calibration Table
No.
Conc.(mg/ml)
1
0.000
C=0.226+1.031*A+^1
Method
Params
8) Calibrate Standard Samples. Put the corresponding standard
samples in the main light path and put Reference in the
reference light path as the screen indicates and press
GOTO ʎ
to set wavelength. Use
EN TER
to set fit method, press
GOTO ʎ
Press
to
,
to select measure method and press
EN TER
10:00:00
D2
W
Abs
WL.(nm)
500.0
r=0.962
Standard
Show Curve
EN TER
to confirm. Input the
EN TER
to confirm.
to confirm.
EN TER
to confirm.
go
into
wavelength
to setup standard, input
to confirm. Users can use
EN TER
to confirm. Press
WL: 500.0nm
Setup Standard Conc.
No.
Conc.(mg/ml)
1
0.000
2
1.000
3
0.000
C=0.226+1.031*A+^1
r=0.962
Input Standard Conc.:0.0000_
Params
,
to
and
,
to
setting
,
CLEAR
to delete,
ES C
to
10:00:00
D2
W
Abs
WL.(nm)
500.0
Select
S TART
S TOP
to