Cleaning And Storage - CareFusion Alaris Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 56
Before the transfer of the pump to a new patient and periodically during the use, clean the Docking Station by wiping over with a lint-free
cloth lightly dampened with warm water and a standard disinfectant / detergent solution. When cleaning the base of the trolley a soft brush
should be used with warm water and a standard disinfectant / detergent solution, due to abrasive finish used on the trolley base.
Recommended cleaners are:
Brand
Hibiscrub
Virkon
Do not use the following disinfectant types:
Disinfectants which are known to be corrosive to metals must not be used, which include:
NaDcc (such as Presept),
Hypochlorites (such as Chlorasol),
Aldehydes (such as Cidex),
Cationic Surfactants (such as Benzalkonium Chloride).
Use of Iodine (such as Betadine) will cause surface discoloration.
Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid to enter the
A
pumps, tiles or electrical contacts and avoid excess fluid build-up on the surface of pumps or tiles. Do not use aggressive
cleaning agents as these may damage the exterior surface of the Docking Station. Do not steam autoclave,
ethylene oxide sterilise or immerse this Docking Station in any fluid.
Information on Disposal for Users of Waste Electrical & Electronic Equipment
U
This
symbol on the product and/or accompanying documents means that used electrical and electronic products should not be mixed
with municipal waste.
If you wish to discard electrical and electronic equipment, please contact your CareFusion affiliate office or distributor for further
information.
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and
the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
This symbol is only valid in the European Union. The product should be disposed of taking environmental factors into consideration. All
components can be safely disposed of as per local regulations.
Maintenance (continued)
Concentration
20% (v/v)
1% (w/v)
1000DF00444 Iss. 2

Cleaning and Storage

Disposal
12/210

Publicidad

Tabla de contenido
loading

Tabla de contenido