The rotating cam can be fitted to the rectangular bar on the Docking Station.
1.
Align the rotating cam on the rear of the pump with the rectangular bar on the Docking Station.
2.
Hold the pump horizontally, push the pump firmly onto the rectangular bar.
Ensure that the pump 'clicks' securely into position onto the bar.
3.
Check that the k on the pump lights up.
The power supply to a tile, and consequently to the pump, is not available until the pump is correctly located on
A
the tile.
4.
To release, push the release lever and pull the pump forwards.
5.
Check that the red LED on the tile is not active once the pump has been removed.
If the LED stays on after removal of the pump, turn the Docking Station OFF, remove from service and contact a
A
qualified service engineer.
Getting Started (continued)
Fitting an Alaris® Infusion Pump
Device
mounting
rail
Release lever (push
to release)
1000DF00444 Iss. 2
Release lever (push to release)
A
It is recommended that infusion bags be located on
a hanger directly above the pump with which they
are being used. This minimises the potential for
confusion of infusion sets when multiple volumetric
pumps are used.
Rotating
cam
8/210
Rotating cam