Mechanical lowering of the actuator
If using the manual/mechanical lowering, please make sure that it will be done in total safety
conditions for the patient. In order to activate the manual/mechanical lowering, pull downwards
the red ring placed in the upper part of the actuator and turn it counterclockwise. Once finished
the lowering process, take the red ring back to its resting position.
Emergency keys
RI830: Underneath the red STOP button of the control box, there is a small button that can be
used to low patients when the remote control fails to operate properly. This button can only be
operated with the aid of a narrow tipped object like the tip of a pen. (Fig.13)
RI835: On the control unit there are two keys (one with an up-facing arrow and the other one with
a down-facing arrow) that can be pressed instead of those ones on the remote control to complete
all the operations when the remote control fails to work properly.
Indication of the battery state of charge (only RI835)
There is a 3-Led indicator for the battery state of charge. This indicator is placed on the remote
control. 3 Leds on indicate a complete charge of the battery, whereas only one Led on indicates
the battery is out of power and needs to be charged.
Recharging batteries
•
Connect the adapter to the main socket and then to the socket on the control unit, after
removing the protection cap.
•
On the charger adapter there is a led that during the charging process becomes orange color.
Charge the battery until the led becomes green color (RI835 only)
•
Before activating the stand up lift again, make sure to disconnect the battery charger!
•
Charge the batteries for at least 6 hours before using the stand up lift.
Fig.13