Philips AVENT SCF334 Manual De Instrucciones página 23

Ocultar thumbs Ver también para AVENT SCF334:
Tabla de contenido

Publicidad

Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctuations/ flicker
emissions IEC
61000-3-3
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
SurgeIEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power
frequency(50/60 Hz)
magnetic fieldIEC
61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment – guidance
Group 1
The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
Complies
network that supplies buildings used for domestic purposes.
IEC 60601 test
level
± 6 kV contact± 8
kV air
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
<5 % UT (>95 %
dip in UT) for 0,5
cycle40 % UT (60
% dip in UT) for 5
cycles70 % UT (30
% dip in UT) for
25 cycles<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Compliance
level
± 6 kV contact± 8
kV air
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
<5 % UT (>95 %
dip in UT) for 0,5
cycle70 % UT (30
% dip in UT) for
25 cycles<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Electromagnetic environment
– guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Mains power quality should be
that of a typical residential or
hospital environment.
Mains power quality should be
that of a typical residential or
hospital environment.
Mains power quality should be
that of a typical residential or
hospital environment. If the user
of the Philips Avent breast pump
requires continued operation
during power mains interruptions,
it is recommended that the Philips
Avent breast pump be powered
from an uninterruptible power
supply or a battery.Interruptible
electric supply can cause the
breast pump to turn off. This is
acceptable as this will not lead to
unacceptable risks.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical residential or hospital
environment.
English
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Avent scf332Avent scd292

Tabla de contenido