7. INSTRUCTIONS FOR USE

• Only use this device for the recommended applications. This device should
only be used under medical supervision.
• Before administering any treatment, you should become acquainted with
the operating procedures for the treatment, as well as the indications,
contraindications, warnings, and precautions. Consult other resources (ie. IFU)
for additional information regarding the application of microneedling therapy.
How to apply BioSheath:
• While wearing non-latex gloves, obtain a single use
BioSheath and ensure the SkinPen
disinfected.
• While SkinPen
Precision between the white tab and paper backing.
• Push SkinPen
device is snug inside the BioSheath.
• Peel back the protective BioSheath cover by pulling on the
Blue tab and white paper backing.
• Remove adhesive backing and seal end. SkinPen
now protected and ready to use.
6
PN#16130015 | Rev. C | International
Precision is clean/
®
Precision is powered off, insert the SkinPen
®
Precision through the BioSheath until the
®
®
How to remove the BioSheath and clean the SkinPen Precision Device:
®
Precision is
Note: Soiled gloves should always be disposed of in a biohazard container. Do
not reuse disposable gloves.
Note: The purpose of a sheath is to provide a covering that helps prevent the
transmission of pathogens from one patient to another. SkinPen Precision is
intended to be used only with provided BioSheath.
• Hold the SkinPen Precision perpendicular to the floor, or with
the cartridge attachment tip pointing downwards. Use one
hand to remove the cartridge and dispose of the cartridge in
a sharps container.
• Continue to hold the SkinPen Precision device perpendicular
to the floor, with the cartridge tip pointed downwards, and
pull apart the adhesive strip of the BioSheath.
• Remove the BioSheath by carefully rolling it down the
SkinPen Precision to prevent soiling the handpiece.
• Dispose of the BioSheath in a biohazard container.
BioSheaths are not intended to be reused.
• Disinfection of the SkinPen Precision should be completed
with the use of Germicidal disposable wipes, See section 9-
Cleaning of SkinPen Precision and Charger Base.
• After removal of the BioSheath and disinfection with
Germicidal disposable wipes is performed, users' gloves
should be removed, hands cleaned, and a new pair of clean
gloves worn before proceeding to the next patient.