SCALER MODULE

Ultrasonic oscillations may prevent cardiac pacemakers functioning properly. Therefore, EMS recommend that
patients with a cardiac pacemaker should not be treated with scaler.
The scaler instruments oscillate at high frequency during the treatment. In rare cases, these oscillations may
cause the instrument to fracture during the treatment. To prevent the patient swallowing or inhaling the fractured
instrument, he must be instructed to breathe through the nose during treatment.
Never use the scaler instruments dry. If used dry, the tips will heat immediately. This may cause thermal injury
to the tooth. Ensure that adequate liquid coolant is always available.

SCALER MODULE WITH FIBRE OPTIC

Ultrasonic oscillations may prevent cardiac pacemakers functioning properly. Therefore, EMS recommend that
patients with a cardiac pacemaker should not be treated with scaler.
The scaler instruments oscillate at high frequency during the treatment. In rare cases, these oscillations may
cause the instrument to fracture during the treatment. To prevent the patient swallowing or inhaling the fractured
instrument, he must be instructed to breathe through the nose during treatment.
Never use the scaler instruments dry. If used dry, the tips will heat immediately. This may cause thermal injury
to the tooth. Ensure that adequate liquid coolant is always available.
When picking the handpiece up, the keys of spray and power
adjustment light up their leds. Notice that spray can only give water,
since air is not used in the scaler.
On the two right side cells of the display the memorized power
appears (between 01% and 100%).
The handpiece can be sterilized according to the indications
given in the table at page 50.
When picking the handpiece up, the keys of spray, fibre optic and
power adjustment light up their leds. Notice that spray can only give
water, since air is not used in the scaler.
The display will show:
a number from 0 to 9 in the first left side cell, indicating the stored
intensity level of the fibre optic,
and a number in percent (between 01% and 100%) in the two right
side cells, corresponding to the stored power intensity.
The handpiece can be sterilized according to the indications
given in the table at page 50.
MODULES
MODULES