SENSOR SETUP

SELECT SENSOR (if there is more than one sensor)
CALIBRATE
Behavior of the outputs during calibration and zero
SET OUTMODE
point adjustment
HOLD
ACTIVE
SET TRANSFER
SENSOR MEASURE
Current, uncorrected measurement value
FACTOR: 0.1 to 100
CONFIGURE
OFFSET: –1000 to +1000
FACTOR
OFFSET
2 POINT
3 POINT
4 POINT
5 POINT
Is shown when FACTOR has been selected under
FACTOR
CONFIGURE. Refer to
detailed information.
Is shown when OFFSET has been selected under
OFFSET
CONFIGURE. Refer to
detailed information.
Is shown when 2 POINT has been selected under
2-POINT
CONFIGURE. Refer to
detailed information.
Is shown when 3 POINT has been selected under
3-POINT
CONFIGURE. Refer to
detailed information.
Is shown when 4 POINT has been selected under
4-POINT
CONFIGURE. Refer to
detailed information.
English 12
Calibration on page 13
for
Calibration on page 13
for
Calibration on page 13
for
Calibration on page 13
for
Calibration on page 13
for
SENSOR SETUP (Continued)
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
Is shown when 5 POINT has been selected under
5-POINT
CONFIGURE. Refer to
detailed information.
SET CAL DEFLT
Security query, reset to (FACTOR=1, OFFSET=0)
CONFIGURE
Name can include up to 16 characters
EDIT NAME
DEFAULT CONFIG: sensor serial number
PAH: Measurement value related to PAH calibration
standard
SET PARAMETER
OIL: Measurement value related to oil calibration
standard
DEFAULT CONFIG: PAH
ppb, ppm, µg/L, mg/L,
SELECT UNITS
DEFAULT CONFIG: ppb
1 to 300 s,
AVERAGE
DEFAULT CONFIG: 3 s
5 s, 30 s, 1, 2, 3, 4, 5, 6, 10, 15, 30 min,
LOG SETUP
DEFAULT CONFIG: 10 min
Range 0 to 500 ppb: AUTO, 0.01 to 50, 0.01 to 500,
Range 0 to 5000 ppb: AUTO, 0.01 to 500, 0.01 to
GAIN VALUE
5000
DEFAULT CONFIG: AUTO
Security query,
SET DEFAULTS
reset to default configuration for all menu options
listed above.
Calibration on page 13