HACH LANGE FP 360 sc Manual Básico Del Usuario página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
Adjustment of slope and zero point; multi-point calibration
Adjust the slope (FACTOR)
1.
Open the MAIN MENU.
2.
Select SENSOR SETUP and confirm.
3.
Select the corresponding sensor and confirm.
4.
Select CALIBRATE and confirm.
5.
Select CONFIGURE and confirm.
6.
Select FACTOR and confirm.
7.
Enter the calculated factor and confirm.
8.
Go back to the MAIN MENU or the Measurement mode display.
Adjust the zero point (OFFSET)
1.
Open the MAIN MENU.
2.
Select SENSOR SETUP and confirm.
3.
Select the corresponding sensor and confirm.
4.
Select CALIBRATE and confirm.
5.
Select CONFIGURE and confirm.
6.
Press OFFSET and confirm.
7.
Enter the required offset and confirm.
8.
Go back to the MAIN MENU or the Measurement mode display.
Multi-point calibration (2 to 5-point calibration)
Note: Multi-point calibration means that the pairs must be input in
ascending order.
1.
Open the MAIN MENU.
2.
Select SENSOR SETUP and confirm.
3.
Select the corresponding sensor and confirm.
4.
Select CALIBRATE and confirm.
5.
Select CONFIGURE and confirm.
6.
Select the type of calibration, for example 2 POINT and confirm.
7.
Select 1PAIR and confirm.
8.
Edit the TARGET VALUE and confirm.
9.
Edit the ACTUAL VALUE and confirm.
10. Repeat the process for 2PAIR and confirm.
11. Go back to the MAIN MENU or the Measurement mode display.
English 15

Publicidad

Tabla de contenido
loading

Tabla de contenido