56
3.2 Input, Output, and Measurement Specifications
3.2 Input, Output, and Measurement
Specifications
-1. Basic specifications
Measurement
Resistance measurement
item
Current measurement
Input
Resistance range:
specifications
Current range:
Range switching: Auto range
Maximum rated
600 V AC (Measurement category IV)
terminal-to-
Anticipated transient overvoltage 8000 V
ground voltage
-2. Measurement specifications/
Accuracy specifications
Common measurement specifications
Accuracy
Accuracy guarantee period: 1 year
guarantee
conditions
Accuracy guarantee period after adjustment made by Hioki: 1 year
Accuracy guarantee temperature and humidity range:
Temperature
(Measurement accuracy × 0.1)/°C is added to the measurement
coefficient
accuracy
(Specified outside a range of 23°C ±5°C)
Nominal
Operating temperature and humidity range:
operating
ranges
Position (Reference position):
Battery voltage:
Earth potential: 3 V rms (Direct current or sinusoidal wave)
External magnetic field:
0.20 Ω/2.00 Ω/20.00 Ω/50.0 Ω/100.0 Ω/200.0 Ω/400 Ω/600 Ω/
1200 Ω/1600 Ω
20.00 mA/200.0 mA/2.000 A/20.00 A/60.0 A
(Clamp sensor opening/closing count: Up to 10000 cycles)
23°C ±5°C, 80% RH or less
See "Operating temperature and humidity range" in "General
Specifications" (p.55).
The measurement target must be placed to pass through the
clamp sensor's center perpendicularly to the clamp sensor.
Available effective battery voltage 2.3±0.19 V to 3.45 V
400 A/m or less
(DC and frequency 50 Hz/60 Hz)
30 A/m or less
(AC current measurement only, frequency 15 Hz to 400 Hz
(excluding 50 Hz/60 Hz))
HIOKI FT6380B983-00