4. Equipment Description

High temperature ovens HIGHTEMP have been designed with fan
assisted air circulation, that allow adjustable temperature from 60ºC up
to 400ºC.
The temperature is controlled by a digital regulator type PID.
As a safety element, a second temperature regulator acts as a
temperature limiter.
The motor can be activated during the cooling cycle, regardless the
he ater.

5. Technical Features

(See also the oven's features plate.)
• Maximum temperature:
• Minimum temperature:
• Volume:
• Fuse size 10x38 gG: (Supply 400V III)
• Fuse size 10x38 gG: (Supply 230V III)
• Max. Power:
• Power Supply:
• Shelf guides:
• Noise level:
• Inner chamber materials:
• Weight:

6. Installation

6.1 Location

You must choose a suitable location for the equipment in order to
avoid accidents and troubles to other laboratory activities, bearing in
mind that this machine produces an unpleasant noise level, and that
the oven generates fumes when operating.
It is important that the location meets the following conditions:
- It should be wide enough to operate with safety.
- There should be a ventilated area, either by windows or extractor
fan, just in case there's a fumes leakage.
- There should be a three-phase + neutral wire to the mains.
- The equipment should be placed over a flat solid surface, suitable for
the equipment weight.
- For a better operation of the equipment, do not expose it near
sunny places nor heating sources.

6.2 Power Supply Connection

Take care when connecting this oven to the mains.
As it has a three-phase supply, the oven motor would turn one-way or
the other depending on how the phases are connected.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - http://www.jpselecta.es
(It can be modified without notice)
Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
Page: 11
400ºC
60ºC
80 litres
Power: 3x 16A
Control: 1x 2A
Power: 3x 20A
Control: 1x 2A
4000W
230 / 400V III 50 Hz
4
70 dB - 80dB
Stainless steel
75 Kg