Task Screen - Selecta Electemp-TFT Manual De Instrucciones

Regulador de procesos
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
CORRECT T
CORRECT. T
(0) 176ºC
ALM OVER T
ALARM SOBRE T
0.0 ºC
ALM UNDER T
ALARM BAJO T
0.0 ºC
Fig 8.2.4 Temperature alarms
Temperature probe
Temperature
reading
set.
100ºC
95ºC
Ramp
2ºC/min
Fig 8.3.1. Run screen operating at one tem-
perature.
J
.P. SELECTA s.a.u.
CODIGO 80080 REV L
8.2.4 Set the temperature and the temperature
alarms
It is recommended to verify calibration of temperature with some regularity
(for example, annually), by using for instance a thermometer or a temperature
pattern.
If it is necessary to correct the reading, as a result of calibration, one must
enter a correction.
To do this, press on this icon:
A box indicating the current correction value appears.
Adjust with keys
temperature pattern.
Applied correction will appear in brackets.
8.2.4.1 Temperature alarms
By pressing successively on the temperature correction box, one may enter an
over-temperature and low-temperature limit. Both alarms activate a message
on screen, but it does not control the process, it's just an alarm.
8.2.5 Temperature units
Press on the Temperature units icon. Press successively to change between
Celsius or Farenheit degrees.
8.2.6 Set language
Press on the Language icon. Press successively on the box to select among
Spanish, French or English.
Adjust by pressing on the Increase and Decrease icons.
8.3 TASK screen
There are two operation modes:
Working at a unique temperature.
Working according to a pre-programmed temperature profile.
Press on
8.3.1 Operation at a temperature
Select the set point temperature by pressing on the temperature indicator.
Select a temperature increase ramp if you want to increase it in a slowly
01:00
way.
01:00
The value "--" makes that the ramp increases as fast as possible (without
ramp control)
Select an operating time. This time starts to count down when the tempe-
rature reaches the set point.
The 0ºC value sets a continuous operation.
Start by pressing on this icon:
The process will end if, the STOP icon is pressed or if it has a programmed
time which comes to an end. When finished, the message "Process completed"
will appear in a text box. Click on the text box to accept the completion.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
MAYO 2019
(Sujetas a modificaciones sin previo aviso) Pag.: 21
 
until you get the indicated value on the
to switch between modes.
Tel 34 937 700 877 Fax 34 937 702 362

Publicidad

Tabla de contenido
loading

Productos relacionados para Selecta Electemp-TFT

Este manual también es adecuado para:

3000897

Tabla de contenido