5. Technical data

Required power supply:
Dimensions:
Weight:
Operating environment
(Operation)
Operating environment
(Transport, storage)

6. Setup and commissioning

6.1 Working principle

The acidified sample solutions are put into the cuvette holder and placed in the premarked
positions of the NANOCOLOR
sample solutions by pressing down the lowering mechanism of the NANOCOLOR
After switching on the device, air is pumped through the sample solutions in the cuvettes
for five minutes. Hereby the inorganic carbon is removed as carbon dioxide quantitatively.
After elapsed time the process stops automatically. The pipette tips are removed from the
solutions and can be disposed using the removing mechanism on top. Afterwards the sample
solutions are treated following the next steps of the respective TOC test kit.

6.2 Package content

®
NANOCOLOR
TIC‑Ex
Cuvette holder
Mains adapter 100–240 V, 50/60 Hz, 9 V
User manual (this leaflet)

6.3 Setting up the instrument

To secure a problem‑free sample purging process, please note the following remarks:
!
Attention:
Place the instrument on a hard and even surface.
Do not move the device during purging process.
Place the device in a dust free environment.
external mains adapter:
Input 100 V–240 V; Output 9 V, 2.0 A
350 mm x 300 mm x 235 mm
4.95 kg (without power supply)
5 °C–40 °C, 20%–80 % relative humidity
5 °C–40 °C, 20%–80 % relative humidity
®
TIC‑Ex. The preinstalled pipette tips are dipped into the
+ 3 adapters
®
NANOCOLOR
TIC-Ex V 2.0 / 11.17
®
TIC‑Ex.
11