Battery Charging; Battery Replacement; Cleaning Instructions - DeVilbiss Healthcare Vacu-Aide 7310 Serie Manual De Instrucciones

Ocultar thumbs Ver también para Vacu-Aide 7310 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
NOTE– Suction ceases if liquid level reaches float shut-off valve located on underside of container lid.
CAUTION– Further suctioning may cause damage to the vacuum pump and voids warranty. Equipment service is required
if fluid content is aspirated back into the unit.
WARNING
This device is rated for intermittent operation use as noted in the specifications. Do not operate device with no
flow or suction port blocked. Continuous operation under this condition may cause the unit to thermal shut off or
damage the device.

BATTERY CHARGING

The Vacu-Aide Compact series is equipped with a factory-installed high capacity rechargeable battery. Connect battery and
charge before first use (refer to Battery Replacement).
1.
Connect the unit to either an AC or DC power source.
2.
The green external power light should illuminate; the yellow charge indicator will remain lit while the battery is
charging.
3.
Ensure that the yellow charging light is illuminated when charging begins. As the battery nears a full charge, the
yellow battery charging light may flash for several minutes. This is normal. If your unit does not hold a charge, check
that the yellow light turns on when external power is applied with the power button "Off." If problems persist, contact
your medical equipment provider.
NOTE– A fully discharged battery will require up to 5 hours (depending on depth of discharge) of charging to reach full
capacity.
If unit is not in use for extended periods, the battery should be recharged a minimum of every 6 months. A fully charged
battery will provide approximately 45-60 minutes of continuous operation at zero vacuum level (free flow).
CAUTION– Completely discharging the battery will shorten the battery life. Do not operate the unit for more than a few
minutes if the low battery indicator light is lit. Recharge the battery as soon as possible.

BATTERY REPLACEMENT

1.
Using a coin or straight-blade screwdriver rotate latch to unlocked position (FIG. 5).
2.
Remove door by pulling up on latch (FIG. 6).
3.
Remove battery from compartment and unplug connector from circuit board (FIG. 7 & 8).
4.
Install new battery by reversing the above steps.
5.
Fully charge battery before using.

CLEANING INSTRUCTIONS

WARNING
To prevent possible risk of infection from contaminated cleaning /disinfection solutions, always prepare fresh
solution for each cleaning cycle and discard solution after each use.
Collection Container:
NOTE– Elbow is permanently attached to the lid and cannot be removed.
1.
Push power button to turn unit off. Wait for vacuum level to drop.
2.
Disconnect external power source from input receptacle on unit (if applicable).
3.
Remove container from unit by disconnecting the patient tubing and filter.
NOTE– The reusable collection container and lid are single-patient use and should be replaced every 1 to 2 months after a
minimum of 30 cleaning cycles as described below.
4.
Wash in a warm water/dishwashing detergent solution and rinse with clean, warm tap water. Then soak with a
commercial disinfectant or one part vinegar (>=5% acetic acid concentration) to three parts warm water
(131ºF-149ºF/55ºC-65ºC) solution for 60 minutes. Thoroughly rinse with warm tap water and air dry. NOTE– If using
a commercial disinfectant, follow manufacturer's recommended instructions and dilution ratios carefully.
NOTE– Elbow is permanently attached to the lid and cannot be removed. Lid/Elbow assembly cannot be autoclaved.
Suction Unit (single-patient use)
1.
With the power "Off," disconnect the unit from all external power sources.
2.
Wipe the outside housing with a clean cloth dampened with commercial disinfectant/detergent (bacterial-germicidal).
CAUTION– Do not submerge suction unit in water; this will result in damage to vacuum pump.
A-7310
EN - 7

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido