CLEANING

There are no routine calibration or adjustment procedures required with the
®
Companion
Enteral Pump. For trouble-free operation, check the pump daily and
clean immediately after spills occur.
While cleaning, pump should be turned off and unplugged. Do not
submerge, autoclave, heat, steam, ETO or radiation sterilize the pump.
THE PUMP CAN BE CLEANED AS FOLLOWS:
HOUSING
Clean outside surface only (DO NOT SPRAY WATER OR CLEANING
SOLUTION INTO PUMP).
For general cleaning, use a soft, lint-free cloth to avoid damaging the surface. Be
especially careful with the operator's panel. If necessary, mild warm, soapy water
may be used sparingly. Avoid the use of all strong solvents or cleaners, as these
will damage the pump.
Disinfection is recommended for all external parts and surfaces. Clean the pump
before disinfecting it. Wipe on the disinfectant for at least one minute; then allow
pump to air dry.
Intermediate level disinfection is achieved by using one of the following: 10%
concentration of 5.25% sodium hypochlorite (household bleach) or 70%
isopropyl alcohol.
NOTE: These recommendations are not substitutes for official procedures that may
differ among institutions.
19