SPECIFICATIONS

POWER
Power:
Fuse:
Power Cord:
Leakage:
MECHANICAL
Height:
Width:
Depth:
Weight:
OPERATIONAL SPECIFICATIONS — FLOW RATES
Range:
Increments:
Accuracy:
Pressure:
Operating Temperature: 60º F – 90º F (16º C – 32º C)
Operating Humidity:
Operating Altitude:
STORAGE
Be sure pump dial is in the OFF position.
Storage Temperature:
Storage Humidity:
Storage Altitude:
120 VAC, 60 Hz, 1 Phase, 6 watts
1/4 AMP, 3AG type
Hospital grade (10 feet), nondetachable from charger
Less than 100 microamps
PUMP ONLY
4.3 inches
6.0 inches
1.7 inches
1.5 pounds
5 to 300 mL/hr
1 mL/hr
±10% with measured flow rates of 1 liter of enteral
product from 5 to 300 mL/hr using Companion
set at zero back pressure (atmospheric).
Pumps against 24 psi (mean) back pressure before
occlusion alarm.
30% RH – 90% RH, non-condensing
The system is capable of successful operation during
exposure to surface elevations from sea level to a pressure
corresponding to 10,000 feet (3,050 meters) above sea
level.
32º F – 95º F (0º C – 35º C)
10% RH – 90% RH, non-condensing
The system is able to withstand and operate after exposure
to elevations up to 20,000 feet (6,100 meters) above sea
level.
21
PUMP WITH CHARGER
6.0 inches
6.7 inches
3.3 inches
4.0 pounds
®