VISUAL DISPLAYS AND AUDIO ALARMS

VISUAL DISPLAYS
CONTROL
DIAL SETTING
POWER OFF
(Charge)
SET RATE
(No pumping action)
RUN
HOLD
(No pumping action)
VOLUME FED
(No pumping action)
VOLUME CLEAR
(No pumping action)
Intermittent audio alarms are always accompanied by a visual message on the display
panel indicating the cause of the alarm. The visual display LOW ALARM indicates that
the alarm volume is set to low. To change the alarm volume, press the HI/LO Alarm
Volume button.
NOTE: If the microprocessor or motor malfunctions, pump will stop pumping
action, all visual displays will cease, and a continuous audio alarm will sound. The
continuous alarm will not cease even if pump dial is turned to HOLD. If this happens,
TURN PUMP DIAL TO POWER OFF; SERVICING IS REQUIRED.
VISUAL DISPLAY
None
Previous flow rate or
000 mL/hr
Numerical rate in mL/hr
and RUN
Existing flow rate or
000 mL/hr Visual alarm
display remains visible (if
one occurred)
Numerical value of
accumulated volume fed
in mL or 0000
Numerical value of
accumulated volume fed
is automatically cleared to
zero (0000 mL)
11
AUDIO ALARM
None
None
None
None
(Any previous audio
alarm ceases)
None
None