Trouble Shooting Guide; Specifications - Mangar International Airflo Mk3 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42
Compressor does not run when 'up' is
selected on the 2-way hand control.
Ensure that the air hose from the product
being inflated is connected to the Airflo.
Compressor keeps stopping when trying to
inflate the lifting aid.
Check that there are no kinks in the air hose.
Battery State Indicator light flashing 'Red/
Green'.
This happens when the current overload
cut-out has tripped. Leave for 3 seconds to
self reset.
Airflo
Weight:
Output pressure:
Battery:
Battery Life:
Lifting performance (
from full battery
Surfer Bather - 10 bathing cycles; Handy Pillow Lift - 11 lifts; Sit-u-Up - 20 lifts; Leglifter - 8 lifts
Environmental protection:
(Airflo housing)
(Push-button hand control) IP67 (
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation.
Charger: Model number - UE10WCP1-150035SPA
Input:
Output:
Double insulated Class II equipment
Simple solutions for everyday independence

Trouble shooting guide

Specifications

4.6Kg (10lbs)
0.35 bar (5 psi)
1 x 12Vdc - 2.3Ah
Maintenance free
(
Approx 1000 charging cycles (
): Bathing Cushion - 4 bathing cycles;
IP51 (
Protection against large dust deposits & vertical falling drops of water
Totally protected against dust & immersion in water
AC 100v - 240v, 50-60Hz, 500mA
DC 15v, 350mA
Battery State Indicator light flashing 'Red'.
This happens when the compressor's
internal temperature cut-out has tripped.
Leave the Airflo for 60-90 minutes to cool.
Compressor causes electrical interference
with another piece of electrical equipment.
Move either the compressor or other item
further away from each other until the
interference stops.
Valve Regulated Sealed Lead Acid)
2 - 5yrs Depending on usage
)
Airflo Mk3
)
)
7

Publicidad

Tabla de contenido
loading