BD Veritor Plus Instrucciones De Uso página 34

Ocultar thumbs Ver también para Veritor Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 145
BD Veritor™ Plus Analyzer Instructions for Use
Result Number
21 / 22
23
24
25
26
27
28
29
* outside of the United States contact your local BD representative
34
Message
The shelf life of the
BD Veritor™ Plus Analyzer
has expired. The maximum
shelf-life of the BD Veritor™
Plus Analyzer from the date
of manufacture is 34 months
ANALYZER
(regardless of time of first
EXPIRED
use).
NO DAYS
REMAIN
The operational life of the
BD Veritor™ Plus Analyzer
has expired. The maximum
number of months from first
use is 24 months.
The operational life of the
ANALYZER
BD Veritor™ Plus Analyzer
EXPIRED
has expired. The maximum
NO TESTS
number of tests performed by
REMAIN
a BD Veritor™ Plus Analyzer
is 10,000 runs.
INVALID BARCODE
ON TEST DEVICE
The test device barcode
REPEAT TEST WITH
cannot be read correctly.
NEW TEST DEVICE
VERIFY FAIL
The result of a BD Veritor™
CONSULT
System Verification Cartridge
MANUAL
run is unsuccessful.
ERROR 25
The BD Veritor™ Plus
RESULT INVALID
Analyzer has detected
FLU A: + FLU B: +
potential interference in the
CONFIRM RESULT
test
ERROR 26
line area.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 27
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 28
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 29
Possible Cause(s)
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
or
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
Outside of the United States contact your local BD
representative
Repeat the assay. On subsequent occurrences, contact
BD*.
Ensure that no foreign matter is present on the
BD Veritor™ System Verification Cartridge assay area,
and repeat the run. On subsequent occurrences, contact
BD*.
Specimens generating a "Result Invalid" with Error 26
should be retested. Upon retesting, if the specimen
produces a "Result Invalid" the user may want to consider
other methods to determine whether the sample is
positive or negative for influenza virus.
Contact BD*.
Contact BD*.
Contact BD*.
Corrective Action(s)

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para BD Veritor Plus

Este manual también es adecuado para:

256066

Tabla de contenido