ENG
8.1
Sterilizing the hand grip
►
Sterilization must be done according to ISO 17665-1
(Sterilization of health care products in moist heat).
CAUTION
Damage to the hand grip
►
Do not sterilize with hot air
►
Package the hand grip in a sterile bag before
sterilization.
►
The hand grip is designed exclusively for damp
sterilization with 3 times fractioned pre-vacuum and
saturated steam with the following parameters:
Temperature
Overpressure
Dwell time
Vacuum drying
►
After sterilization, check the hand grip for mechanical
integrity.
►
Do not use damaged hand grips.
9.
SAFETY INSPECTIONS
DANGER
Death hazard from electric shock.
►
Unplug from the grid and turn switch to off position.
►
The power supply cable must be checked at least
once a year for damage.
CAUTION
►
Maintenance and repairs can be performed only by
qualified electricians.
►
The corresponding user profile is in Chapter 2 Safety
instructions.
EVERY YEAR:
►
Check the power supply cable for damage and
replace if necessary.
►
Check for deformations and cracks in the plastic parts.
►
Check the load bearing system for distortion or
damage.
►
Check for loose parts.
36
9.1
Triango 80 C / Triango 80 W
134°C
2.0 bar
6 min
20 min
►
►
►
►
SETTING THE SPRING FORCE
CAUTION
►
The ideal spring force is set at the factory.
CAUTION
►
The luminaire head must be installed before the
spring force is adjusted.
Move the spring balanced arm to the highest possible
position.
Insert a 6 mm Allen wrench through the opening on the
middle joint and adjust the spring force by turning the
spring.
Turn the screw clockwise (+) to increase the spring force
(if the spring balanced arm lowers).
Turn the screws counterclockwise (-) to reduce the
spring force (if the spring balanced arm rises).