OHAUS CKW Serie Instrucciones De Instalación Y Operación página 7

Ocultar thumbs Ver también para CKW Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN-6
Menu Settings (Cont.)
Only commands shown in Table 1, Serial Interface Command Table, are
acknowledged by the indicator. The default serial output format from the RS485
interface is shown as follows:
F
ield:
@
Address
Length:
1
Definitions: "@":
Address: A 2 digit number from 00 to 99.
Polarity:
Weight:
Units:
Stability: "?" character is printed if not stable, blank if stable.
Reset
Used to reset the Communication menu to factory
defaults. Selections are: No, Yes.
Baud Rate
Used to define Baud rate. Selections are:
300, 600, 1200, 2400, 4800, 9600 or 19200. . . . .
Parity
Used to define parity. Selections are:
None, Odd or Even.
Data Length
Used to define data length. Selections are:
7 or 8.
Stop Bit
Used define the number of stop bits. Selections are: 1 or 2.
Polarity
Space
2
1
1
The "@" character and 2 digit address are automatically
added when a RS485 address is enabled.
"-" sign if negative, blank if positive.
Up to 6 digits and 1 decimal, leading zero blanking.
Up to 5 characters.
CKW Serial Interface Option Kit
Weight
Space
Units
7
1
Stability CR
5
1
LF
1
1

Publicidad

Tabla de contenido
loading

Tabla de contenido