9. When the status
indicator light stays
blue, remove the
calibration lid.
13. Install the
calibration lid. Make
sure that the calibration
lid is in the closed
position. Refer to
Installation
on page 3.
17. Remove the vial.
12 English
10. Remove the vial.
14. Push the button on
the front of the
instrument.
18. Make sure that
there is no water on the
process head (or the
automatic cleaning
module). Dry all
possible spills to
prevent water ingress
on the vial
compartment.
11. Put the 600 NTU
vial in front of the RFID
module. A beep sound
is heard and the status
indicator light flashes
blue. If the status
indicator light does not
flashes blue, refer to
Troubleshooting
on page 22.
The instrument records
the value, the lot
number, the expiration
date and the Certificate
of Analysis information
from the RFID vial to
the data log.
15. Wait 30 to
60 seconds for the
measurement to
complete. The status
indicator light slowly
flashes blue during the
measurement.
19. Hold the process
head (or automatic
cleaning module)
vertically when it is
installed on the
instrument or the vial
can break.
12. Put the 600 NTU
vial in the vial
compartment.
16. When the status
indicator light flashes
green, remove the
calibration lid.
20. Push the button on
the front of the
instrument to save the
calibration value. The
status indicator light
stays green.