Section 1 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
1.1 Product overview
The calibration lid is used with the TU5300 sc and TU5400 sc turbidimeters for calibration and
calibration verification with sealed StablCal vials or user-prepared formazin.
1.2 Product components
Make sure that all components have been received. Refer to
damaged, contact the manufacturer or a sales representative immediately.
Figure 1 Product components
1 Calibration lid
Section 2 Installation
Keep water out of the vial compartment or instrument damage will occur. Before the calibration lid is installed on
the instrument, make sure that there are no water leaks. Make sure that all tubing is fully seated. Make sure that
the vial nut is tight.
Hold the process head (or automatic cleaning module) vertically when it is removed from the instrument or
condensation water can fall into the instrument. If condensation water gets into the vial compartment instrument
damage will occur.
Make sure to lift the process head (or automatic cleaning module) a sufficient distance to release the vial
(approximately 10 cm (3.94 in.) or the vial can break. If the vial breaks, water will get in the vial compartment and
instrument damage will occur.
Do not to touch or scratch the glass of the process vial. Contamination or scratches on the glass can cause
measurement errors.
Note: Make sure that no particles fall into the vial compartment.