Donning The Device; Fitting Advice - Blatchford CTTCP22L Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42

9 Donning the Device

Care must be taken during donning/doffing not to damage the device with fingernails,
sharp jewelry or the locking pin.
Do NOT pull or stretch the device.
1.
Roll the device inside out.
2.
Make sure the device is positioned
correctly to suit the top profile.
3.
Roll the device on taking care to expel any
trapped air.
After donning the device, allow the user to wear it
for 10 minutes. If the user feels numbness, tingling
or any unusual sensation within 10 minutes, do the
following:
4.
Doff the device, and wait until normal sensation returns.
5.
Don the device again.
6.
If the user feels numbness, tingling or any unusual sensation again, stop using the
device.

10 Fitting Advice

Cushion
The device must be worn in conjunction with an air-
tight sleeve.
Locking
When using a locking liner the integrated matrix
should finish just below fibula head/patella tendon
level.
The locking pin must always follow the long axis of
the residual limb. (See diagrams.)
The pin should be secured by applying
Loctite 243 to the threads of the locking pin.
Torque the locking pin to 3 Nm, or tighten it with
your fingers by either 1/8 to 1/4 of a turn.
Locking Pin
Shoulder
8
938335PK1/1-0521

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cttcp40lCttlp22lCttlp40l

Tabla de contenido