The Versatile Table installation must be done by Olsen certified technician consists in assembling legs
rest and checking the mains voltage which will supply the equipment and, if necessary, make the tension
adjustment. In the case of optional items have been purchased, these should also be installed at this time.
The technician will also verify if the equipment complies with the purchasing order, if it maintains its
integrity, and will provide operation, cleaning and maintenance guidance.
When the installation is complete, observe the following items:
− Both right and left bracket locks are installed with the upholstered leg rest;
− All remote foot pedal commands work perfectly;
− The sliding extender has the stainless-steel tray and the upholstery and slides perfectly;
− The upholsteries of backrest, seat, headrest and legs rest are intact.
In the case of optional items have been purchased, check the items below according to the acquired
optional item:
− The armrests are installed on the side rails;
− The Colposcope holder was installed and performs all the joints perfectly;
− The knee supports were installed on their respective bracket locks;
− The heel supports were installed on their respective bracket locks;
− Integrity of upholstery and finishes;
− Joints without rigidity when moving;
− The fitting of accessory is secure.
For instructions on the proper functioning of the equipment and its components, read Chapter 5
(Description and Operating of the Equipment) of this manual.
Installation should only be made by an authorized technician. Installation performed by an
unauthorized person will result in loss of warranty.
To position the equipment, put the backrest parallel to the floor and pull the sliding extender and the
retractable step until your limit, and then bring the equipment to where it will be used. Carry out the up and
down movements of the backrest as well as open and storing the sliding extender and the retractable step,
making sure that there is enough space for all movements.
To access the Olsen certified technical assistance network for installation and maintenance contact us by
e-mail
export3@olsen.odo.br
The whole sanitize process must be done with the use of gloves suitable for cleaning and
protection, in addition to a mask and protective glasses, according to biosafety standards.
The upholsteries, plastic covers and laminated parts must be cleaned with a damp cloth containing
neutral soap or detergent only. Olsen advises against using any chemical product to clean these parts, but in the
case of disinfection products, it is important to check if it has suitable compatibility and specifications before
use on these materials.
Never use hypochlorite or alcohol-based products.
It must be cleaned with a slightly damp cloth containing only soap or mild detergent.
Never use hypochlorite or alcohol-based products.

8 - Installation

or if you prefer +55 48 2106 6000.

9 - Cleaning and Disinfection