Follow the instructions in chapter 7 of this manual (Installation Requirements) to suit the electrical
network where the equipment will be installed.
Follow instructions for proper use of the equipment and its accessories as described in chapter 5
(Description and Operating of the Equipment). Improper use could be harmful to equipment, which
would not be covered by warranty.
Follow the instructions in chapter 9 (Cleaning and Disinfection) of this manual for daily cleaning of
your equipment.
Protect your equipment from direct exposure to sunlight. The equipment direct exposure to sunlight
may cause premature aging of plastic covers and upholstery.
Turn off the circuit breaker or unplug the equipment at the end of the working day.
In case of remote foot pedal damage, discontinue use of the equipment, turn it off, and contact
Olsen accredited assistance.
The electric mains cable was developed for use only on the Versatile Table. The use of this
component in other equipment may compromise the emissions and electromagnetic immunity of
these.
Only the authorized technician can replace the equipment´s mains cable and internal fuses.
This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air, O
or Nitrous Oxide.
This equipment should only be operated by physicians and nurses for clinical, outpatient, and
gynecological procedures.
Use only the mains cable supplied with the equipment for mains connection. Use of cables other
than those specified (chapter 10 - Technical Features) may result in increased emissions or reduced
immunity of the Versatile Table.
Do not remove the plastic covers from the equipment. The risk of electric shock! Only the accredited
technician is authorized to perform this procedure.
In the case of damage to the electrical panel, adjacent plastic covers and motors plastic covers
disconnect the equipment from the mains and contact the Olsen accredited service. The use of the
equipment must be interrupted until maintenance is complete. The use of the equipment in these
conditions offers a risk of electric shock!
Do not perform maintenance or cleaning on the Versatile Table while it is in use or turned On.
Do not install or use any electrical equipment over or near the Versatile Table. If necessary, the
Versatile Table must be checked to see if it is functioning normally in the configuration in which it
will be used.
Do not perform fuse replacement if it is possible to touch the patient, even if unintentionally, during
the procedure.
20