Screw the control valve firmly onto the cylinder and connect the gas
supply as shown below:
Control
Valve
20 ppm
carbon
monoxide
in air
The plastic tubing supplied with the gas should be pushed firmly over
the reducing connector.
Plastic Tubing
Flow
Indicator
Reducing Connector
for Calibration
51