For the purposes of EN 60601-1-2 the Smoke Check monitor has an essential
performance (the Smoke Check should continue to indicate a level of 11-20
p.p.m. during a test, whilst measuring gas from a 15 p.p.m. CO gas bottle)
Guidance and Manufacturer's Declaration – Electromagnetic Emission
The Smoke Check is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or the user of the Smoke Check should
ensure that it is used in such an environment
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Smoke Check is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or the user of the Smoke Check should
ensure that it is used in such an environment
Immunity Test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient / burst
IEC61000-4-4
Surge
IEC61000-4-5
Compliance
Group 1
Group B
Not Applicable
(battery powered)
Not Applicable
(battery powered)
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
Electromagnetic Environment -
Guidance
The Smoke Check uses RF energy
only for its internal function. Therefore,
its RF emission is very low and is not
likely to cause any interference in
nearby electronic equipment.
The Smoke Check is suitable for use in
all establishments, including domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes
Compliance
Electromagnetic
Level
Environment - Guidance
± 6 kV contact
Floors should be wood,
± 8 kV air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Not Applicable
Mains power quality
(Battery
should be that of a typical
powered)
commercial or hospital
and no cables
environment.
Not Applicable
Mains power quality should
(Battery
be that of a typical
Powered) and
commercial or hospital
no cables.
environment.
59