a least once per year at a minimum but depending on the application and risk potential of the application a greater
frequency might be needed. Check with local authorities for any specific local regulations.
Contents of the Cal-Kit 1
Cal-Kit 1
(30-0011-1110-2)
o
Calibration Case
o Two feet of Tygon tubing
o
Cal Hood-Macurco Pack
o 0.2 LPM Gas regulator (F)
Also needed are the following gas bottles depending on the target gas (Sold Separately):
Hydrogen
o 70-0716-1758-6 Hydrogen H2 Cal Gas Cylinder 34L 10% LEL (M) (For Gas Testing)
o 70-0716-1759-4 Hydrogen H2 Cal Gas Cylinder 34L 20% LEL (M) (For Calibrating)
Methane
o 70-0716-1754-5 Methane CH4 Cal Gas Cylinder 34L 10% LEL (M) (For Gas Testing)
o 70-0716-1755-2 Methane CH4 Cal Gas Cylinder 34L 20% LEL (M) (For Calibrating)
Propane
o 70-0716-1756-0 Propane C3H8 Cal Gas Cylinder 34L 10% LEL (M) (For Gas Testing)
o 70-0716-1757-8 Propane C2H8 Cal Gas Cylinder 34L 20% LEL (M) (For Calibrating)
Cal-Kit 1 Information
Several detectors can be calibrated with one Cal-Kit. The only limitation is the amount of gas in the cylinder. The 34-
liter cylinder has approximately 170 minutes of continuous calibration run time. The gas cylinder should be replaced
when the pressure gauge on the regulator shows 25-psi or less.
Note: For optimum test results it is suggested that the unit be in clean air, green light on, and be in a low ambient air
flow.
REV – 1.0.0
[34-2900-0027-6 ]