Programming
Press Fn for 2 seconds in the start screen to program the device with 6
modes (PED1 - PED6).
Press OK to pass from a parameter to the next one.
It is possible to change the duration of all programs from 5' to 100' with
steps of 5'. Default duration 20'.
PED1
It is possible to set pulse frequency and duration.
From 1 to 150 Hz with steps of 1 Hz.
From 50 to 250 µs with steps of 10 µs.
Default values: 100 Hz / 200 µs.
PED2
It will be possible to set pulse duration with a fixed frequency of 100 Hz.
From 50 to 250 µs with steps of 10 µs.
Default values: 200 µs.
PED3
It is possible to set frequency values.
From 1 to 150 Hz with steps of 1 Hz.
Default values: 100 Hz
Modulated pulse duration from 100 µs to 250 µs.
PED4
It is possible to set pulse frequency and duration.
From 1 to 150 Hz with steps of 1 Hz.
From 50 to 320 µs with steps of 10 µs.
It is possible to change the following values: tUP = 1 s; tHOLD = 5
s; tDOWN = 1 s; tOFF = 5 s.
Default values: 30 Hz / 250 µs.
tUP
tHOLD
tDOWN
tOFF
PED5
It is possible to set frequency values.
From 1 to 150 Hz with steps of 1 Hz.
Default values: 30 Hz
Modulated pulse duration from 100 µs to 320 µs.
51
Rev. 03 Multi_06_2018