6.4.3 - Scaler Adjustments - Operation and Basic Adjustments

To activate the scaler: push the propulsion/activation
pedal (6).
To adjust the vibration power: turn the control knob (7)
until the desired power is reached.
To toggle functions: push the control knob (7) in
direction of the table for "GP" function (General - Perio). Pull the
knob gently in the opposite direction of the table for "E" (Endo)
function.
To adjust the water: use the control button (8) on the instrument control panel.

6.4.4 - Scaler Tips

The tips GD1, PD1 and PD3 accompany the scaler. The power and application information for each tip is
available in the table below.
Note:
1 - Do not use the tip on features other than those specified by the manufacturer.
2 - Do not exceed the power limit for each tip. That could damage the tip and the scaler.

- Curing Light (Optional)

Before starting the curing light use, install the
tip (1) and the light hood (2).
To install the tip: first, attach the light hood (2)
to the tip and then attach it to the curing light
unit (3). The tip must be pushed all the way to the fitting end.
To activate/deactivate the curing light: press the On/Off button (5) twice. The
green LED (4) indicates that the curing light is operating. The curing light unit will
run for 20 seconds and will turn Off automatically.
To toggle the operating mode: each tap on the selection button (9) switches
the operation mode. Press the selection button (9) until the orange LED indicates
the desired mode.
The curing light operates in the following modes:
− Continuous (6): turns on the maximum power for 20 seconds;
− Ramp (7): turns on the minimum power, gradually increasing. In 5 seconds
reaches its maximum power until the cycle of 20 seconds is finished;
− Pulse (8): the device blinking for 20 seconds.

- 3-Way Syringe

Before beginning the use of the 3-way syringe (5), connect the
syringe tip (3) by pressing the locking ring (4) into the correct fit.
To emit air jet: press the air button (1).
To emit water jet: press the water button (2).
To emit spray jet: press simultaneously the air and water
buttons (1 and 2).