Pro-care Optima / English
4. After completing initial inflation, system will conduct automatic detection of patient weight
and provide corresponding pressure to optimize mattress pressure relieving performance.
NOTE- Patient is NOT supposed to lie on the mattress before initial inflation is completed,
but to be on the mattress when the auto adjustment process about to start.
NOTE- Once the patient is on the mattress, the pump will automatically sense and adjust
the pressure in the cells to support the patient.
NOTE- Pressing the knob for a minimum of 3 seconds can quickly start the auto
adjustment.
5. Main LCD screen displays the current therapy. Setting can be changed by turning the knob and
the mode which is chosen will be highlighted, then press the knob once to enter or confirm
the setting.
Mode
5.1 Mode
Alternating
NOTE- When changing between operating modes, the patient's monitoring and
repositioning program should be reviewed.
5.2 Cycle time
10 mins
5.3 Tuning
Original setting
The 5 blue bars displayed on the screen represents the status of the
automatic adjustment detection. When completes, the 5 bars will stop
running and become
Cycle time
Continuous Low Pressure
15 mins
Decrease
mattress pressure
.
Tuning
Back to previous page
25 mins
20 mins
Increase
mattress pressure
14
Back to previous page
Back to previous page