How to define the patient tolerance • At the start of a session
This function, MODULATION Mode is a way to set within the tolerance of a patient at the
beginning of a session.
Set up the patient and proceed as below:
Beginning
Switch the unit ON
To select the MODULATION Mode
To select the pain level
(
for flexion,
for extension)
To set the pain level when reached,
immediately press
To start the session with the new movement
limits
Comments: When the maximum or minimum range of motion of the machine is reached, this
limit is recorded and the MODULATION Mode is automatically switched OFF.
How to define the patient tolerance • During the session
This function, The BY-PASS MODE is a way to address the pain threshold of a patient during a
session, which allows you to work on increasing amplitude.
Beginning
The unit is running
To select the BYPASS mode
To select the NEW pain level
To set the NEW pain level when reached,
immediately press
Continue the session with the new movement
limits.
Comments: When the maximum or minimum range of motion of the machine is reached, this
limit is recorded and the BYPASS Mode is automatically switched OFF.
Spectra & Spectra DC
Keys to press
Or
Continuous press
Or
IMPORTANT: Can be used only when the machine RUN.
Keys to press
Or
Continuous press
Or
Display
KINETEC
40 STOP
50 110
MODUL. : use +or-
40 STOP
50 110
MODUL. : use +or-
40 MANUAL 60
110
MODUL. : use +or-
40 MANUAL
60
60
KINETEC
40 STOP
55
60
Display
KINETEC
40 RUN
55
60
WARM UP
40 RUN
55
60
BYPASS : use +or-
40 RUN
55
60
BYPASS : use +or-
40 BYPASS
72
60
BYPASS : use +or-
40 BYPASS
72 72
KINETEC
40 RUN
64
72
EN
Remarks
Check if the hand control is not
blocked
(See page 5).
If the Warm Up mode is selected,
switch OFF this mode by pressing on
the key «Warm Up».
The display indicates the keys to be
used to drive the machine, this
message is displayed 3 seconds.
The unit is moving to the selected
way.
It is waiting for you to select new
limits.
The new limit of the movement is
recorded.
To start the session with the new
movement limits.
Remarks
Check if the hand control is not
blocked
(See page 5).
If the Warm Up mode is selected,
switch OFF this mode by pressing on
the key «Warm Up».
The display indicates the keys used to
run the machine.
This message is displayed 3 seconds.
The unit is moving to the selected
way.
It is waiting for you to select new
limits.
The new limit of the movement is
recorded.
The angle display changes with
current movement.