Foundation Fieldbus:
Resource Block and
Transducer Block
Value Error
0
Other
1
Block Configuration Error
2
Link Configuration Error
3
Simulate Active
4
Local Override
5
Device Fault State Set
6
Device Needs Mainte-
nance Soon
7
Sensor Failure
8
Output Failure
9
Memory Failure
10
Lost Static Data
11
Lost NV Data
12
Readback Check Failed
13
Device Needs Mainte-
nance Now
14
Power-up
14
Out-of-Service
Transducer Block
Value Error
16
Unspecified error
17
General error
18
Calibration error
19
Configuration error
20
Electronics Failure
21
Mechanical Failure
22
I/O Failure
23
Data Integrity Error
24
Software Error
25
Algorithm Error
BLOCK_ERR: Este parámetro de bloque de error refleja el estado de error de los
componentes asociados de hardware o software y los impactos directos sobre la operación
correcta de un bloque. Los bloques de error se notifican como alarmas de bloque.
Description
For the resource block, Simulate Active will be used to indicate that the simulate hardware
jumper is present. An active state (1) of this attribute will indicate that the jumper is present
and that it is possible for the user to enable simulation in an input or output class function
block.
detected by this block/process variable has a status of BAD, Sensor Failure
detected by this block/back calculation input has a status of BAD, Device Failure
XD_ERROR: Un parámetro adicional , XD_ERROR, tiene que usarse para localizar los
subcódigos de error individuales que el operario considere más importantes cuando
ocurren uno o más errores. BLOCK_ERR tendrá establecido un bit 0 siempre que
XD_ERROR no sea cero.
Description
An error has occurred that was not identified.
An error has occurred that could not be classified as one of the errors below.
An error occurred during calibration of the device or a calibration error has been detected
during operation of the device.
An error occurred during configuration of the device or a configuration error has been
detected during operation of the device.
An electronic component has failed.
A mechanical component has failed.
An I/O failure has occurred
Indicates that data stored within the system may no longer be valid due to NVM checksum
failure, Data verify after write failure, etc.
The software has detected an error. This could be caused by an improper interrupt service
routine, an arithmetic overflow, a watchdog timer, etc.
The algorithm used in the transducer block produced an error. This could be due to an
overflow, data reasonableness failure, etc.
Polytron 7000 funcionamiento con PROFIBUS PA
85