Operation
Replacing the Fuses
WARNING:
To avoid electric shock, remove test leads from the meter
before removing the fuse door.
1. Remove the two Phillips screws from
the battery/fuse door located on the
back of the meter.
2. Remove door.
3. Remove the old fuse from its holder
by gently pulling it out.
4. Install the new fuse into the holder.
5. Always use a fuse of the proper size
and value (UL rated 0.2A /600V fast
blow for the 200mA range, UL rated
10A /600V fast blow for the 10A range).
6. Install battery/fuse door and tighten
screws.
WARNING:
To avoid electric shock, do not operate your meter until
the fuse door is in place and fastened securely.
FUSE
9V BATTERY
FUSE
11
Specifications
Function
Range
200mV
DC Voltage
2000mV
20V
200V
600V
200V
AC Voltage
600V
2000µA
DC Current
20mA
200mA
10A
200
Resistance
2000
2Ok
200k
2000k
Battery Test
9V
1.5 V
o
NOTE: Accuracy is stated at 65
F to 83
Resolution
Accuracy (% of reading)
0.1mV
1mV
±(0.5% reading + 2 digits)
0.01V
0.1V
±(0.8% reading + 2 digits)
1V
0.1V
±(1.2% reading + 10 digits)
1V
50/60Hz
1µA
±(1.0% reading + 2 digits)
10µA
100µA
±(1.2% reading + 2 digits)
10mA
±(2.0% reading + 2 digits)
0.1
1
±(0.8% reading + 4 digits)
0.01k
0.1k
1k
±(1.0% reading + 2 digits)
10mV
±(1.0% reading + 2 digits)
10mV
±(1.0% reading + 2 digits)
o
o
o
F (18
C to 28
C) and less than 75% RH.
12