in place when the jaw of stapler closes.
3. Hold and squeeze the trigger to be in the semi-closure status, and now the tissue in
the jaw can be adjusted.
4. After adjusting the tissue, squeeze the trigger tightly until the jaw of stapler closes.
If relax the grip, the trigger will fully rebound to the original position.
5. Fully squeeze the trigger to stitch and cut the tissue, and then continuously
squeeze the trigger until the trigger firing does not move, which means that the
triggering process finishes.
6. Press the blue reset button at the end of the stapler to open the jaw of the stapler,
and then pull the button of locating pin back to the original position.
Note: After moving away the stapler, check the bleeding conditions of suture lines,
and if there is mild bleeding, manual suture or electric coagulation hemostasis can be
applied.
[Component replacement method]
1. Before removing the components of the used stapler cartridge, make sure that the
jaw of the stapler is in the fully open status, and the locating pin has been fully
returned back.