2. Pinch the thumb pad beyond the component of the stapler cartridge, then draw it
upward to draw out the jaw. Before replacing the components of stapler cartridge,
any tissue and/or shaping but still unused suturing nail shall be removed away from
the surface of the cutting stapler device, which can ensure the correct shaping of
suturing nails when triggered.
3. Replacement of components: Pinch the thumb pad of the components of the
unused stapler cartridge, and make the pin hole align with the cutting stapler device.
Insert the components of the stapler cartridge into the clamping plate of metal
components, and then push down until the components of the stapler cartridge are
in the proper position.
Note: Before installing the components of the stapler cartridge, check whether the
button of locating pin is in the initial position, and then reset the button of locating pin
to the initial position, if not, for normal installation of stapler cartridge components; if
the components are not installed in the proper position, or fail to be fully inserted, or
fully/partially triggered, the jaw of the stapler may not close normally. Please select
the proper components of stapler cartridge based on the thickness of the tissue.
4. Repeat the above-mentioned steps based on the needs.
Note: Each stapler can be replaced with 7 components of the stapler cartridge for
eight times of triggering.
5. Suture and cutting operations can be done again according to the above methods
from Step 1-4.
STORE AT ROOM TEMPERATURE.
AVOID PROLONGED EXPOSURE TO ELEVATED TEMPERATURES.
DO NOT EXPOSE TO TEMPERATURES ABOVE 1300F (54
From the EO sterilization date, the product is valid for 3 years.
℃) .