... fortuitous reading.
Broken cable.
Substitute the electrode.
... impossible to perform calibration.
... very slow response.
Dirt on the diaphragm and/or the membrane.
De-hydrated membrane.
Electrode ageing or wear.
Clean, regenerate or replace the electrode.
Specifications of 50 54 T electrode
pH measuring range
Operating temp.
Reference element
Diaphragms
Electrolyte
Body material
Min. immersion depth
Applications:
• Cheeses, meats, fi sh, fruits, bread doughs, etc. It can be used for
measurements in aqueous solutions too.
Limits
pH values <2.
Temperatures > 60 ºC.
Warranty
CRISON guarantees this electrode against manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the electrode.
Ordering information
Code
Description
Electrode/dagger complete (composed of 50 54
50 54 TC
puncture electrode and knife 50 54 KN)
50 54
T
Puncture electrode
50 54 KN
Stainless steel knife
QUALITY CERTIFICATE
The attached electrode 50 54 T has passed the quality test and fol-
lows the next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s
• Temperature error <= 0.5 ºC
2 ... 14
0 ... 60 °C
encapsulated Ag/AgCl crystals
1 ceramic + 1 open
solid
glass / 316 s. steel
15 mm
with Pt 1000 A.T.C.
50 54
T
(spare part)
T
(spare part)
11