blackening etc.
4.2.3
Mirror
The presence of impurities or smudges on the tip mirror may lead to
poor scan quality and an overall poor scanning experience. In such a
situation, you should clean the mirror following the steps below:
▪......
Disconnect the scanner tip from the i700 handpiece.
▪......
... P our alcohol on a clean cloth or cotton-tipped swab and wipe
the mirror. Make sure to use alcohol that is free of impurities or
it may stain the mirror. You can use either ethanol or propanol
(ethyl-/propyl alcohol).
▪......
Wipe the mirror dry using a dry, lint-free cloth.
▪......
... M ake sure the mirror is free of dust and fibers. Repeat the
cleaning process as necessary.
4.2.4
Handpiece
After treatment, clean and disinfect all other surfaces of the handpiece
except for the scanner front (optical window) and end (air vent hole).
Cleaning and disinfecting must be done with the device turned off.
Use the device only after it is completely dry.
Recommended cleaning and disinfecting solution:
Denatured alcohol (aka. ethyl alcohol or ethanol) – typically 60-70%
Alc/Vol.
The general cleaning and disinfecting procedures are as follows:
▪......
Turn off the device using the power button.
▪......
Unplug all the cables from the power hub.
▪......
Attach the handpiece cover to the front of the scanner.
▪......
Pour the disinfectant onto a soft, lint-free, and non-abrasive cloth.
▪......
Wipe the scanner surface with the cloth.
▪......
Dry the surface with a clean, dry, lint-free, and non-abrasive cloth.
CAUTION
▪......
... D o not clean the handpiece when the device is turned on as the
fluid may enter the scanner and cause malfunction.
▪......
Use the device after it is completely dry.
CAUTION
▪......
... C hemical cracks may appear if improper cleaning and
disinfecting solutions are used during cleaning.
4.2.5
Other Components
▪......
... P our the cleaning and disinfecting solution onto a soft, lint-free,
and non-abrasive cloth.
▪......
... W ipe the component surface with the cloth.
▪......
... D ry the surface with a clean, dry, lint-free, and non-abrasive
cloth.
CAUTION
▪......
... C hemical cracks may appear if an improper cleaning solution is
used during cleaning.
Intraoral Scanner i700
13