Measuring Instructions - Testboy TV 218 Instrucciones De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
Technical specifications

Measuring instructions

|
For optimal measuring precision: position the cable precisely
at the point inside the clamping jaws where the marker lines
intersect.
|
If the cable is positioned improperly, the measurement error
will be no greater than 1.5%.
Accuracy:  (% of measured reading + number of digits) at
|
18 °C to 28 °C (64 °F to 74°F) and relative humidity < 75%.
AC current measurements (True RMS)
Range
40 A
200 A
DC current measurements
Range
40 A
200 A
(Max: 300 ADC)
34
Resolution
0.01 A
0.1 A
Resolution
0.01 A
0.1 A
Tolerance
± 3.0 % +5 digits
Tolerance
± 2.5 % +8 digits
®
Testboy