Single Point Calibration; Points Calibration - Metria M21 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Single point calibration

1.1
Ensure that the meter is in the pH measurement mode and you have selected 1 point calibration in the setup menu.
1.2
Press the Cal key, the display shows
1.3
Rinse the pH electrode with distilled water, place the electrode (and temperature probe) into the pH7.00 (or 6.86) buffer
solution and stir gently.
1.4
Press the Enter key, the meter begins the calibration, the Calibration icon continuously flashing.
1.5
Wait for the reading to stabilize, the meter will automatically show

2 points calibration

2.1
Ensure that you have selected 2 points calibration in the setup menu.
2.2
Repeat steps 1.2 through 1.4 above. When the first calibration point is completed, the display will show
prompts you
to continue with second point calibration.
2.3
Rinse the pH electrode with distilled water, place the electrode (and temperature probe) into the next buffer solution and
stir gently (E.g.,
pH4.01).
9
Calibration
2.4
Press the Enter key, the meter automatically recognizes the pH buffer solution and begins the calibration.
2.5
Wait for the reading to stabilize, the display will show electrode slope and
3 points calibration
3.1
Ensure that you have selected 3 points calibration in the setup menu.
3.2
Repeat steps 1.2 through 1.4 above. When the first calibration point is completed, the display will show
meter prompts
you to continue with second point calibration.
3.3
Rinse the pH electrode with distilled water, place the electrode (and temperature probe) into the pH4.01 buffer solution
and stir gently.
3.4
Press the Enter key, the meter begins the calibration, the Calibration icon continuously flashing.
/
or
/
, depending on the pH buffer group you selected.
and returns to the measurement mode.
. Calibration is completed.
. The meter
/
. The

Hide quick links:

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido