Ph Calibration; Setting The Number Of Calibration Points - Metria M21 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Preparing the pH buffer solution
1.
Open the pH7.00 buffer sachet, place the reagent into a 250ml volumetric flask.
2.
Fill the volumetric flask to the mark with distilled water, mix the solution until the reagent is completely dissolved.
3.
Preparation of pH4.01 and 10.01 buffer solutions are the same as above. Prepared standard buffer solutions should be
stored in hermetically
sealed glass containers.
8
Calibration

pH Calibration

The METRIA M21 pH meter allows 1 to 3 points calibration in the pH mode. We recommend that you perform at least 2 points
calibration for high accuracy measurement. The meter will automatically recognize and calibrate to following standard buffer
values.
USA Standard Buffers
NIST Standard Buffers
Single point calibration should only be carried out with pH7.00 or 6.86, otherwise calibration will not be accepted.
Make sure to calibrate the meter when attaching a new electrode. Do not reuse the calibration solution after calibration,
contaminants in solution will affect the calibration and eventually the accuracy of the measurement.
In order to get accurate measuring results, we recommend using a stirrer to create the homogeneous buffer solutions and sample.

Setting the number of calibration points

1.
Press and hold the
2.
Press the Enter key, the display shows
Press the  /  key to select 1 or 2 or 3 points calibration, press the Enter key until the meter returns to the measurement
3.
mode.
pH4.01, 7.00, 10.01
pH4.01, 6.86, 9.18
key for 3 seconds to enter the setup menu.
.
Glass
sensitive
membrane
3M KCL solution

Hide quick links:

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido