Table 2
Guidance and manufacture's declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
power supply lines:
transient/burst
±2 kV
IEC 61000-4-4
input/output lines:
±1 kV
Surge
line(s) to line(s): ±1 kV
IEC 61000-4-5
line(s) to earth: ±2 kV
100 kHz repetition
frequency
Voltage dips, short
0%UT; 0.5 cycle
interruptions and
At 0°, 45°, 90°, 135°,
voltage variations on
180°,225°,270° and
power supply input
315°
lines
0%UT ; 1 cycle
IEC 61000-4-11
and
70%UT ; 25/30 cycles
Single phase: at 0°
0% UT ; 300 cycle
(50Hz/60Hz)
30 A/m
magnetic field Power
50Hz/60Hz
frequency
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
Compliance level
±8 kV contact
±15 kV air
power supply lines:
±2 kV
line(s) to line(s): ±1 kV
100 kHz repetition
frequency
0% UT ; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
0% UT ; 1 cycle
and
70% UT ; 25/30 cycles
Single phase: at 0°
0% UT ;300 cycle
30 A/m
50Hz/60Hz
19 I UK
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.