to EU-Type Examination Certificate No. FM15ATEX0034X
Description of Equipment or Protective System:
13
The YCO16-Z Ex-Link Converter is a dedicated device providing power and data to the
PMA.Vision VIS1X...-... weighing platforms (FM15ATEX0032X).
PMA.Vision are available in RS232 communications. The YCO16-Z Ex-Link Converter is housed
within non-metallic enclosure. A dedicated potential equalization connection is provided on both
the PMA.Vision and the YCO16-Z ExLink Converter. Power and data connections between the
YCO16-Z and the PMA.Vision are made via a multi-conductor CAT6 cable.
Electrical ratings
Um = 250 V
Specific Conditions of Use:
14
1. The YCO16-Z Ex-Link Converter shall only be used with the PMA.Vision (FM15ATEX0032X).
2. The protective ground connectors on the PMA.Vision and the YCO16-Z Ex-Link Converter shall be
connected together using a potential equalisation conductor with a cross-sectional area of at least 4
mm2.
3. The CAT6 cable used between the PMA.Vision and the YCO16-Z Ex-Link Converter shall be obtained
from Sartorius.
15
Essential Health and Safety Requirements:
The relevant EHSRs that have not been addressed by the standards listed in this certificate have been
identified and assessed in the confidential report identified in item 8.
16
Test and Assessment Procedure and Conditions:
This EU-Type Examination Certificate is the result of testing of a sample of the product submitted, in
accordance with the provisions of the relevant specific standard(s), and assessment of supporting
documentation. It does not imply an assessment of the whole production.
Whilst this certificate may be used in support of a manufacturer's claim for CE Marking, FM Approvals
Europe Ltd accepts no responsibility for the compliance of the equipment against all applicable Directives
in all applications.
This Certificate has been issued in accordance with FM Approvals Europe Ltd's ATEX Certification
Scheme.
17
Schedule Drawings
A list of the significant parts of the technical documentation is annexed to this certificate and a copy has
been kept by the Notified Body.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Europe Ltd. One Georges Quay Plaza, Dublin. Ireland. D02 E440
T: +353 (0) 1761 4200 E-mail:
atex@fmapprovals.com www.fmapprovals.com
F ATEX 020 (Mar/2019)
SCHEDULE
Appendix
All signal outputs to the
Page 2 of 3
Installation Instructions VIS1X
91

Publicidad

Tabla de contenido
loading

Tabla de contenido