3. Product Description

3.1
Component groups and functioning elements
1 Mixing tube
2 Sensor
3 Gas valve
4 Knurled nut
5 Hose spout
6 Actuation knob for ignition safeguard
7 Collection pan (only 934 0100 and 935 0100)
8 Sensor holder
9 Clamping screw
10 Allen key
8
1
7
3
934 0100 / 935 0100
3.2
Supplied in Delivery
1 Safety Burner eco
1 Instructions for Use
1 Collection pan (only 934 0100 and 935 0100)
1 Allen key

4. Initial Operation

The initial operation of the burner may only be carried out by qualified personnel according to DVGW
stipulations or valid national regulations.
Please observe the following:
⇒ Check if the type of gas and the connection pressure of the burner correspond to the gas supply.
⇒ Check the fit of the connection pipe and that it is not damaged.
⇒ Close the gas shut-off valve at the laboratory workbench (gas isolation valve).
⇒ Slide the certified gas hose (according to DIN 30664 or DVGW G 5501-P) up to the last groove of the
hose spout (5) on the burner. If necessary, secure the hose in position with a clamp or ring clasp (see
DVGW worksheet G 621 or relevant national specifications).
2
9
4
6
5
10
8
1
3
6
935 0200
- 3 -
2
9
4
5
Fig. 1